Directive and declaration of manufacturer – Electromagnetic immunity
OMRON M24/7 is suitable for use in the speci¿ ed electromagnetic environment. The
purchaser or user of OMRON M24/7 should assure that it is used in an electromagnetic
environment as described below.
Immunity test IEC60601-1-2 Compliance
level
Electromagnetic
environment
Conducted RF
IEC 6100-4-6
3V
eff
150KHz - 80MHz
Not applicable Portable and mobile RF
communications equipment
are used no closer to any part
of OMRON M24/7, including
cables, than the Recommended
Separation Distance calculated the
formula written below.
Recommended Separation
distance:
d=[3.5/V
1
]¥P
Radiated RF
IEC 61000-4-3
3V/m
80MHz – 2.5GHz
3V/m d=[3.5/3V/m]¥P;
(80MHz – 800MHz)
d=[7/3V/m]¥P;
(800MHz – 2.5GHz)
where:
P is the highest radiated power
disclosed by the manufacturer of
transmitter [W];
d is the recommended separation
distance [m].
1. note: in case of frequency 80MHz or 800 MHz, the formula for the higher range is
applicable.
2. note: These are guidelines. Actual conditions may vary.