13
A successful monitoring session consists of the following steps:
1. Connecting the OMRON M24/7 monitor to the PC (First connection only
– not required to repeat with each session)
1. Connect the optical interface, which was delivered with the device, to the
USB port of the PC.
2. The other side of the optical USB interface cable, which is a two-point
plastic connection, should be connected to the socket of the OMRON
M24/7, in a way that the red plastic ring is directed to the lower side of the
device.
3. Start the OMRON BP Tracker Software and open the Options menu, then
click on the Communication tab on the left side of the screen.
4. On the appearing window, click on the USB option. Please connect the
device to the PC (fresh/charged batteries, device switched on) as noted
above then click on the Test button. Upon successful communication the
software will display the serial number and the ¿ rmware version of the
device.
2. Preparation of Monitoring Session
1. Inform your patient about the monitoring rules well in advance.
2. Programming the measurement plan into the device:
• Using PC: start the OMRON BP Tracker Software program
• Optionally the device can be programmed without using a PC, for
details see Manual Programming section.
3. Enter new patient data or select patient from the database.
4. Create a monitoring plan with respect to the patient’s lifestyle.
5. Insert two fresh or fully charged, AA size batteries into the battery
compartment and check their voltage.
6. Connect the monitor to the computer.
7. Send the monitoring plan from the computer to the monitor unit.
8. Apply the cuff to the patient with the device placed in the carry pouch.
9. Give the Patient’s Quick Guide to the patient along with detailed
instructions about the rules and the usage of the device.
3. Evaluation
1. Remove the unit and cuff from the patient on his/her return.
2. Ask for the Patient’s Quick Guide /patient diary, and ask the patient
for any events, symptoms, observations or complaints.
3. Start the OMRON BP Tracker Software
4. Connect the device to the PC and then transfer collected data from
the monitor to your database.
5. Analyze the blood pressure pro¿ le.
6. Create and print a report.
IM_EN.indd 13 19/07/2012 14:30:02