4. Handling Errors and Problems
The date and time flash,
even though the date
and time have been set,
after the batteries are
replaced.
The emergency deflation
switch was pressed.
Set the date and time
again. (Refer to Chapter
2.3.)
Even if the power is
turned off, the clock will
also stop if you press the
emergency deflation
switch. Take care not to
press the emergency
deflation switch when
cleaning or moving the
unit.
Have the batteries been
removed?
Set the date and time
again. (Refer to Chapter
2.3.)
The date and time will be
reset if the batteries are
removed for more than
30 seconds. Replace the
batteries within 30
seconds.
No display appears
when O/I button is
pressed.
The batteries are empty.
Replace the batteries.
(Refer to Chapter 2.1.)
The batteries are
installed incorrectly.
Insert the batteries
correctly.
(Refer to Chapter 2.1.)
The AC adapter was
disconnected from the
unit when using the AC
adapter only.
Connect the AC adapter.
(Refer to Chapter 2.2.)
The AC adapter was
disconnected from the
power supply when using
the AC adapter only.
Connect the AC adapter
to the power supply.
(Refer to Chapter 2.2.)
Readings are not stored
in the memory.
G (guest) was selected,
or no user ID was
selected.
Select the correct user
ID. (Refer to Chapter
3.3.)
Problem Cause Remedy