Maintenance
Maintenance Inspection and Safety Management
The HBP-1300 must be maintained to ensure functionality and to secure the safety of patients and
operators.
Daily checks and maintenance should be performed by the operator. (page 25)
In addition, qualified personnel are necessary to maintain the performance and the safety, and to
conduct periodic inspections. We recommend that the verification test be performed at least once a
year. (page 17)
Managing Supplies
Keep the following supplies on hand.
GS CUFF - All sizes
Device Maintenance
Cleaning and disinfecting should be performed in accordance with your facility's infection control
practice and OSHA regulations.
Surface cleaning
Wipe with a cloth that has been moistened with isopropyl alcohol diluted to 50 v/v%, or ethyl alcohol
(disinfection alcohol) diluted to 76.9 - 81.4 v/v% and wrung out.
Do not wipe the DC jack or allow it to become wet.
Removing dust
Use a moistened cotton bud to remove dust that has accumulated on the vent ports.
Service
The device requires no routine service other than cleaning, and visually checking the cuffs, tubing, etc.
Caution
Do not sterilize by autoclave or gas sterilization (EOG, formaldehyde gas, high-concentration ozone,
etc.).
If using an antiseptic solution for cleaning, follow the instructions of the manufacturer.
Accessory Care
Non-Invasive Blood Pressure Measurement (NIBP)
Cuff
Wipe clean on the surface of the cuff with a cloth moistened with a 70 v/v% dilution of isopropyl alcohol,
or a 76.9 to 81.4 v/v% dilution of disinfection ethanol (ethyl alcohol).
Do not allow any liquids inside the cuff. If a liquid gets in the cuff, dry the inside well.