Omron HBP-1300 Blood Pressure Monitor User Manual


 
19
Non-Invasive Blood Pressure (NIBP) Measurement
Cuff Selection and Connection
Selecting the cuff
Caution
Do not use the cuff if it is damaged or has holes.
Use the appropriate cuff size to ensure correct measurements. If too large a cuff is used, the
measured blood pressure value tends to be lower than the actual blood pressure. If too small a cuff is
used, the measured blood pressure value tends to be higher.
Note:
It is important to use the correct sized cuff for a patient in order to get an accurate reading.
Measure the circumference of the patient's arm and select the cuff size
that is appropriate for the circumference.
Select the cuff that is suitable for the patient from the cuffs below.
* Available as an optional accessory.
Connecting the cuff
Connect the cuff tube to the NIBP connector on the unit and turn
clockwise to lock.
Caution
Only OMRON GS CUFF can be used with this device.
Note:
If the cuff tube is too short, the optional 3.5 m or 1.5 m CUFF HOSE can be connected to lengthen the
cuff tube. Do not make the hose longer than 3.5 m, as this will affect measurement accuracy.
Make sure the connections are tight.
Warning
Treat a cuff used on a patient with an infection as medical waste, or sufficiently sterilize the cuff
before reuse.
Cuff name
Arm circumference
(inch) (cm)
GS CUFF XL (GCUFF-XL)* 17-20 42 - 50
GS CUFF L (GCUFF-L) 13-17 32 - 42
GS CUFF M (GCUFF-M) 9-13 22 - 32
GS CUFF S (GCUFF-S)* 7-9 17 - 22
GS CUFF SS (GCUFF-SS)* 5-7 12 - 18
HBP-1300.book Page 19 Wednesday, March 7, 2012 9:28 AM