System Settings
Utility Mode
"Utility Mode" allows you to configure the "Auto Power Off" and "Pressure Accuracy Confirmation" setting.
1.Confirm that the device is switched off.
2.Hold down the [MODE] button until the
"Auto Power Off" setting screen appears.
The "Initial Inflation Pressure Value" setting screen appears,
and changes to the "Auto Power Off" setting screen appears.
3.Press the [ ] or [ ] button to change the auto power off setting.
The "5 min." or "10 min." setting is entered.
When the unit is powered by the battery pack and is not used for the set time ("5 min." or "10 min."),
the power automatically turns off to save battery power.
If an alarm has occurred, the power does not turn off automatically.
4. When you have completed the "Auto Power Off" setting, press the [MODE] button.
The "Pressure Accuracy Confirmation" screen appears. "0" appears.
5.Check the Pressure Accuracy.
Change the pressure externally.
Compare with the displayed value and verify that there is no problem.
6.When you have completed "Pressure Accuracy Confirmation", press the
[START/STOP] button.
The power turns off.
Example:
1.Connect the blood pressure monitor, the standard pressure gauge, and
the cuff and rubber ball.
For details on the standard pressure gauge, contact OMRON HEALTHCARE Customer Service.
2.Check the blood pressure value of the blood pressure monitor and the
blood pressure value of the standard pressure gauge.
Note:
• The standard is within ± 3 mmHg or 2 % of manometer reading.
• To rapidly deflate the cuff, press the [ ] button.
To repeat "Pressure Accuracy Confirmation", turn off the power and repeat the procedure from step 1
of "Utility Mode".
"5" or "10" appears.
"0" appears.