Omron 705IT Blood Pressure Monitor User Manual


 
English
19
18
Maintenance
and storage
Technical data
For cleaning the monitor only use a soft, lightly
moistened cloth.
Do not use petroleum spirits, thinners or
similar solvents !
Stains on the cuff can be carefully removed using
a moist cloth and soapy water.
Do not wash the cuff or make it wet !
Keep your blood pressure monitor in the storage
case to protect it from dust and moisture.
Protect your monitor against vibrations and do not
leave it in a place where temperatures are
extremely low (below -20°C) or extremely high
(above 60°C) or with high humidity (+ 85%).
Do not use your monitor at very low temperatures
(below 10°C) or very high temperatures (above
40°C).
Do not fold the cuff or air tube together too tightly
and do not crease them.
Do not carry out repairs of any kind yourself.
If a defect occurs, consult the OMRON distributor
or Customer Services or seek the advice of your
surgical appliance stockist or pharmacist.
The accuracy of this blood pressure monitor is
designed to last a long time.
In order to increase the lifetime of the batteries
take them out when you store the device for a
longer period than 2 months.
Electro-magnetic fields can influence the
measurement. Do not use a portable phone near
this unit.
This device fulfils the provisions of the EC
directive 93/42/EEC (Medical Device Directive).
This blood pressure monitor is designed
according to the European standard EN1060,
Non-invasive sphygmomanometers Part 1:
General requirements and Part 3: Supplementary
requirements for electromechanical blood
pressure measuring systems.
Model: OMRON 705IT
Display: LCD digital display
Blood Pressure Pressure: 0 mm Hg to 299 mm Hg
Measurement Range: Pulse: 40 – 180/min.
Accuracy: Pressure: ±3 mm Hg
Pulse: ±5% of reading
Memory: 28 measurements
Inflation: Fuzzy-Logic controlled by electric pump
Deflation: Automatic pressure release valve
Pressure Detection: Capacitive pressure sensor
Measurement Method: Oscillometric method
Power Source: 4 Alkaline batteries 1.5V (Type LR6) or
AC/DC adapter (optional, 6V = 4W)
Battery Life: Capacity of new batteries is approx. 300
measurements
Operating Temperature: 10°C to 40°C
Storage Temperature: -20°C to 60°C
Humidity: 15 to 90% RH maximum
Storage humidity: 10 to 95% RH maximum
Console Weight: Approximately 380g without batteries
Outer Dimensions: Approximately 115
(l) mm x 177 (w) mm x 71 (h) mm
Cuff Dimensions: Approximately 140 mm x 480 mm
(arm circumference 22 – 32 cm)
Accessories: Medium cuff, CD-ROM, USB cable, instruction
manual, storage case, battery set
NOTE: Subject to technical modification without prior notice
= Type B
• Disposal of this product and used batteries should be carried out in accordance
with the national regulations for the disposal of electronic products.