MediSense ART06986 Blood Glucose Meter User Manual


 
Applying Control Solution to
the Test Strip
1
Turn the control solution
bottle upside down three
to four times to mix
the solution.
2
Remove the cap.
3
Apply a drop of control
solution to the test strip in
the area shown here. The
control solution is drawn
into the test strip.
4
Continue to touch the
control solution to the test
strip until the monitor
begins the test.
The monitor begins the
test when:
•You hear the beeper, if the
beeper is ON.
•The display window shows
the status bar.
•Then the display window
shows the countdown.
Glucose
ß-Ketone
Status
bar
Countdown
61
Control Solution
Testing