Lumiscope 2220 Thermometer User Manual


 
7
Measuring Temperature
Toggle between forehead scan mode and object scan mode by
pressing, holding momentarily, and releasing the MODE button.
Taking a temperature reading from the forehead:
Press the POWER button and hold it
for one second to turn the device on.
Release it, and the system will power
up, which includes a self test. During
this period the device will momentarily
display all symbols on the LCD (see
Figure 4).
Press and hold the SCAN button and aim the thermom-
eter at the patient’s forehead, maintaining a distance
of 6 cm ~ 8 cm (2
3
/8" ~ 3
1
/8") between the Non-Touch
Thermometer and the patient’s forehead. You will hear
a continuous beeping sound, accompanied by a ash-
ing LED. Release the SCAN button; you will hear a short
beep. Do not move the thermometer away from the fore-
head until you hear the beep. Moving the thermometer
away from the forehead before hearing the beep may
result in an inaccurate temperature reading. Please note:
you will not hear a continuous beeping sound unless the
thermometer is the correct distance from the forehead
(see Figure 5). Results will display on the LCD.