Invacare TPO100 Oxygen Equipment User Manual


 
SECTION 6—OPERATING INSTRUCTIONS
Invacare SOLO
2
24 Part No 1156872
Connecting the Humidifier
(ifsoprescribedandonlywithContinuous FlowMode)
ƽ WARNING
DONOTuseahumidifier whileinPulseMode.
DONOTov erfill humidifier.
DONOTreversetheoxygeninputandoutputconnections.
Waterfromthehumidifierbottlewilltravelthroughthe
cannulabacktothepatient.
NOTE:Forthisprocedure,refertoFIGURE6.3andFIGURE6.4onpage 25.
NOTE:Connectthehumidifieronlyifprescribed.Usethehumidifieronlyin
ContinuousFlow Mode.
1. Removecapfrombottle.
2. Fillhumidifierwithdistil ledwatertothelevelindicatedbythe
manufacturer.Replacethehumidifiercapand securelytighten.
FIGURE 6.3 Filling the Humidifier
3. Attach
thehumidifierbott leadaptertothehumidifierbottlebyturning
thewingnutonthehumidifierbott lecounterclockwiseuntilitis
securelyfastened.SeeDETA IL“A inFIGURE6.4.
4. Placethehumidifierbottle/adapterassemblyintheoutsidepocketof
theaccessorybag.Attachtheaccessorybagtothepullcartas
shownin
FIGURE6.4.
5. Rotatethebottlewithinthepocketsothatthehumidifieradapterhose
isaw ayfrom thebodyofthetransportableoxygenconcentrator.
6. Attachoxygentubingfromthehumidifierbottle/adapterassemblyto
oxygenoutletconnectorontheconcentrator. SeeDETAIL“B”in
FIGURE6.4.
7. Attachthecannula/patientsupplytubing
tothe humidifierbottleoutlet.
SeeDETA IL“B”inFIGURE6.4.
8. Afterassembly,ensurethatoxygenisflowingthroughthecannula.
Humidifier Bottle
w/o Cap
Humidifier Bottle
with Cap