Invacare L-4 Mobility Aid User Manual


 
14 TROUBLESHOOTING
Part No. 1143205
53 Invacare® Lynx™ L-3 & Invacare® Lynx™ L-4 Scooters
Scooter will not drive. Error
Brake release lever disengaged.
Batteries require charging.
Key not fully plugged in.
Charger plugged in.
Circuit breaker tripped.
Fuse blown in battery box.
Soft stop engaged. If key is turned off while
driving, the scooter holds power for 1.5 seconds
slowly decelerating to a stop.
Battery connections loose.
Check service indicator light on control panel (FIGURE6.1 on
page 25). If flashing, count the number of flashes and refer to
Service Indicator Error Codes
on page 54.
Engage brake release lever. Refer to Engaging/Disengaging the
Brake Release Lever on page 22.
Charge batteries. Refer to Charging the Batteries on page 32.
Ensure key is inserted fully into keyswitch.
Unplug charger from scooter.
Reset breaker. Refer to Resetting the Circuit Breaker
on
page 55. If breaker trips again, it may indicate need for internal
repairs. Contact Dealer or a qualified technician.
Contact Dealer or qualified technician.
With key in the off position, wait for 30 seconds and turn key
to the on position.
Check all connections. Secure connections. Refer to
Removing/Installing the Battery Box
on page 30.
SYMPTOM PROBABLE CAUSE SOLUTION