Invacare IRC5PAW Oxygen Equipment User Manual


 
SECTION 8—TROUBLESHOOTING GUIDE
Perfecto
2
™Series 28 Part No 1150713
YELLOW or RED
light Illuminated.
Alarm:
Continuous
On RED light
only.
* Only applies to
IRC5PO2AW
models.
1. Low oxygen purity.* 1. Clean or Replace filters.
2. Kinked or blocked
tubing, cannula or
humidifier.*
2. Inspect for kinks or
blockages. Correct, clean or
replace item. Once
corrected, turn power Off for
60 seconds and then turn
power back On.
3. Flowmeter set at 0.5
L/min *.
3. Check flowmeter is set to
1.0 L/min or more. Refer to
Typical Product Parameters
on page 13.
4. Unit overheating due
to blocked air intake.
4a. Remove and clean cabinet
filters.
4b. Move concentrator at
least three inches from walls,
draperies, and furniture.
5. Insufficient power at
outlet.
5. DO NOT use extension
cords. Move to another
electrical outlet or circuit.
6. Internal repairs
required.
6. Call Service Provider.
GREEN light with
YELLOW Light
Flashing.
1. Internal repairs
required.
1. Call Service Provider.
Alarm:
Rapid
Beep....Beep...
Beep...Beep
LOW FLOW ALARM
1a. Kinked or blocked
tubing, cannula or
humidifier.
1b. Flowmeter set at 0.5
L/min.
1a. Inspect for kinks or
blockages. Correct, clean or
replace item. Once
corrected, turn power Off for
60 seconds and then turn
power back ON.
1b. Flowrates less than 1
L/min are not recommend.
SYMPTOM: PROBABLE CAUSE: SOLUTION: