SECTION 5—TYPICAL PRODUCT PARAMETERS
Part No 1150713 15 Perfecto
2
™Series
No AP/APG Not suitable for use in the presence of a
flammable anaesthetic mixture.
Standards and Regulatory
Listing:
IRC5PO2AW/IRC5PAW
IEC/EN61000-3-2
IEC/EN61000-3-3,
IEC/EN 60601-1, A1, A2
IEC/EN 60601-1-2
IRC5PO2AW Models Only
ISO8359
MDD 93/42/EEC, Annex I and IX
CE marked model IRC5PO2AW
Electrical: No extension cords.
Placement: No closer than 7.5 cm (3 in) from any wall,
furniture, draperies, or furniture to assure
sufficient air flow. Avoid deep pile carpets and
heaters, radiators or hot air registers. Floor
location only.
Tubing: 2 m (7 ft) cannula with a maximum 15 m (50 ft)
of Crush-Proof Tubing (DO NOT pinch).
Time of Operation: Up to 24 hours per day.
Recommended Storage
and Shipping Temperature:
-29°C to 65°C (-20°F to 150°F) at 15-95%
relative humidity.
Environment: Smoke and soot-free. No confined spaces
(Example: No closets).
Minimum Operating Time: 30 Minutes