Invacare IRC5PAW Oxygen Equipment User Manual


 
SECTION 7—MAINTENANCE
Perfecto
2
™Series 24 Part No 1150713
SECTION 7—MAINTENANCE
ƽ WARNING
TheInvacareconcentratorsarespecificallydesignedto
minimizeroutinepreventivemaintenance.Onlyqualified
personnelshouldperformpreventivemaintenanceonthe
concentrator.
Unplugtheconcentratorwhencleaning.Toavoidelectrical
shock,DONOTremovecabinet.
NOTE:Ataminimum,preventivemaintenanceMUSTbe
performedaccordingtothemaintenancerecordguidelines.Inplaces
withhighdustorsootlevels,maintenancemayneedtobeperformed
moreoften.RefertoPreventiveMaintenanceRecordonpage 26.
Routine Maintenance
Thefollowingroutinemaintenanceshouldbeperformedmore
frequently.Seerecommendedintervalsineachsectionand
PreventiveMaintenanceRecordonpage 26.
Cleaning the Cabinet Filter
CAUTION
DONOToperatetheconcentratorwithoutthefilterinstalled.
NOTE:Forthisprocedure,refertoFIGURE7.1onpage 25.
NOTE:Thereisonecabinetfilterlocatedonthebackofthecabinet.
1. Inspectthefiltermonthly.Removethefilterandcleanas
needed.
NOTE:Environmentalconditionsthatmayrequiremorefrequent
inspectionandcleaningofthefilterinclude,butarenotlimitedto:
highdust,airpollutants,etc.