Invacare IRC5LXO2AWQ Oxygen Equipment User Manual


 
SECTION 6—MAINTENANCE
Platinum™ 30 Part No 1134866
FIGURE 6.2 Preventive Maintenance Record
ON EACH INSPECTION
MODEL NO. IRC SERIAL NO.
PREVENTIVE MAINTENANCE RECORD
INVACARE PLATINUM OXYGEN CONCENTRATOR
Record Date of Service
Record Elapsed Hours On Hour Meter
Clean Cabinet Filter(s)
Check Prescribed L/min. Flow Rate
EVERY4,380 HOURS FOR IRC5LXAW
Check Oxygen Concentration
(UNITS WITHOUT SensO
2
)
Clean/Replace Cabinet Filter(s)
Check Outlet HEPA Filter*
Check Compressor Inlet Filter*
(
Check Power Loss Alarm
AS REQUIRED
Clean Heat Exchanger
Replace Exhaust Muffler
Rebuild Top End of Compressor
*NOTE: Refer to Preventive Maintenance section of Service Manual.
NOTE:
2,160 hours are equivalent to usage 24 hours per day, for 90 days.
4,380 hours are equivalent to usage 24 hours per day, 7 days per week, for 6 months.
26,280 hours are equivalent to usage 24 hours per day, 7 days per week, for 3 years.
Check Oxygen Concentration
UNITS WITH SensO
2
)
DURING PREVENTIVE MAINTENANCE SCHEDULE, OR BETWEEN PATIENTS
EVERY 26,280 HOURS OR 3 YEARS, WHICHEVER COMES FIRST