Invacare IRC5LXO2AWQ Oxygen Equipment User Manual


 
SECTION 5—OPERATING INSTRUCTIONS
Platinum™ 26 Part No 1134866
FIGURE 5.7 SensO
2
Oxygen Purity Indicators (LX O
2
Sensor
Model)
GREENlight(O
2
)‐NormalOperation.
YELLOWlight()‐CallsupplierIMMEDIATELY.Youmay
continuetousetheconcentratorunlessinstructedotherwise
byyoursupplier.Becertainthatbackupoxygenisnearby.
REDlight()‐Totalunitshutdown.SwitchIMMEDIATELY
tobackupoxygensupply.CallsupplierIMMEDIATELY
(FIGURE5.7).
GREENlight‐withYELLOWlightflashing‐Callsupplier
IMMEDIATELY.Oxygensensormalfunctioning;youmay
continuetousetheconcentrator(FIGURE5.7).
Indicators (LX Model)
NOTE:Forthisprocedure,refertoFIGURE5.8.
Explanation of LX models Indicator Lights
REDlight()‐TotalUnitShutDown.Switch
IMMEDIATELYtoabackupoxygensupply.Callsupplier
IMMEDIATELY.
GREENlight(I/O)‐On/Off.Systemokay.
O
2
Auto Shut Down
Use Backup
Below
Normal
Normal
RED
Call Supplier
YELLOW
GREEN
OXYGEN PURITY