Invacare IRC5LXO2AWQ Oxygen Equipment User Manual


 
SECTION 5—OPERATING INSTRUCTIONS
Platinum™ 20 Part No 1134866
Recommended Guidelines for Optimum
Performance
NOTE:Ensurethatyourconcentratorisatleast7.5cm(3in)away
fromwalls,draperiesorfurnituretoassuresufficientairflow.Avoid
deeppilecarpetsandheaters,radiatorsorhotairregisters.
Plug in Power Cord
1. Pluginpowercordtoanelectricaloutlet.
Temperature: (10°C - 35°C) 50°F - 95°F
Electrical: No extension cords.
Placement: No closer than 7.5 cm (3 in) from the wall,
furniture, draperies, or similar surfaces.
Tubing and Cannula: IRC5LXAW, IRC5LXO2AW,
IRC5LXO2AWQ
2.1 m (7 ft) cannula with a maximum 15.2 m
(50 ft) of Crush-Proof Tubing (DO NOT
pinch).
IRC9LXO2AWQ
Recommended use up to 15.2 m (50 ft) high
flow tubing with high flow cannula at all flow
rates.
Environment: Smoke and soot-free. No confined spaces
(Example: No closets).
Relative Humidity: 20 to 60%
Time of Operation: Up to 24 hours per day.
Flow Rate IRC5LXAW, IRC5LXO2AW,
IRC5LXO2AWQ
0.5 L/min. to 5 L/min. Flow rates less than 1
L/min., are not recommended.
IRC9LXO2AWQ
1 L/min. to 9L/min. Flow rates less than 1
L/min. are not recommended.
Minimum Operating Time: 30 Minutes