Inova CTPT-6LPZ Oxygen Equipment User Manual


 
718
Indications for Use
The SmartDose Liquid Oxygen Portable CTPT-3LPZ and/or
CTPT-6LPZ provides a source of supplemental oxygen for
ambulatory home healthcare patients, by vaporizing medical
grade liquid oxygen and then dispensing it to the patient via an
integral electronic conserving device. The SmartDose Liquid
Oxygen Portable CTPT-3LPZ and/or CTPT-6LPZ is neither a
life-sustaining nor a life-supporting device.
P
rovider
C
hecklist
Before leaving the SmartDose Liquid Oxygen Portable with a
patient, the following checklist MUST be completed:
Check all parts for shipping damage. In case of damage, do not
use. Contact CHAD Therapeutics for further instructions.
Make sure the SmartDose Liquid Oxygen Portable is compatible
with the liquid oxygen source. Be sure to use a liquid oxygen
reservoir with a compatible
99% coupler.
Instruct the patient on the safe operation of the SmartDose Liquid
Oxygen Portable and all associated accessories and review ALL
warnings.
Leave a copy of this product manual with the patient.
T
ransporting and Storing SmartDose Liquid Oxygen Portables
Preparing for Shipping
Prior to shipping the SmartDose Liquid Oxygen Portable, remove the
batteries and empty any residual oxygen from the unit. Allow the
portable to rest for at least 8 hours to empty.
Storage
Store the SmartDose Liquid Oxygen Portable in a cool, clean, dry area
when not in use. Avoid storing in direct sunlight.
Battery Disposal
Dispose of all batteries in accordance with local, state, and federal
regulations.
Spec
ifications
Batteries Required* 2 AA batteries
*If using rechargeable batteries, it is recommended to use >
1800 mAh
Ni-MH (Nickel Metal Hydride) batteries. Do NOT mix battery types.
Oxygen Pressure 22 to 25 psig
CTPT-3LPZ Content Volume 0.33 liters
(approx)
CTPT-6LPZ Content Volume 0.6 liters (approx)
CTPT-3LPZ Weight (empty) 3.5 lb
CTPT-6LPZ Weight
(empty) 4.5 lb
Storage Temp/Transport -20 to 50ºC
Operating Temperature -8 to 40ºC
Storage Humidity 0 to 95% RH
(non-condensing)
Operating Humidity 15 to 95% RH