IDEAL INDUSTRIES Meter Thermometer User Manual


 
Page 5
DC CURRENT:
DCA 61-633 61-635
40mA, 400mA, 4A, 10A ± (0.50% + 4d) ±(0.20% + 4d)
AC CURRENT:
ACA 61-633 61-635
40mA, 400mA, 4A, 10A ±(1.20% + 8d) ± (0.80% + 8d)
Bandwidth 40Hz ~ 400Hz 40Hz ~ 400Hz
Range: 40mA, 400mA, 4A, 10A
Resolution: 1µA in the 40mA range
Burden Voltage: 800mV max. for mA input, 1V max. for A input
AC Conversion Type: AC Coupled True RMS responding
Input Protection: Equipped with High Energy Fuse
1A, 600V, IR 10KV fuse (Bussmann BBS-1 or equivalent) for mA input
15A, 600V, IR 100KV fuse (Bussmann KTK 15 or equivalent) for A input
AC + DC Current: Same as AC(RMS) + 1.00% + 8d
C.F.: Same as ACV
PEAK HOLD: +[± (0.7% + 20)] additional error for >10% of full scale
RESISTANCE:
OHM 61-633 61-635
4000, 4K 0 ± (0.50% + 2d) ± (0.30% + 2d)
40KQ, 400K0 ± (0.50% + 2d) ± (0.30% + 2d)
4M0 ± (0.50% + 4d) ± (0.30% + 4d)
40M0 ± (5.00% + 5d) ± (5.00% + 5d)
LV OHM 61-633 61-635
4K0 ,40K0 400K 0 ± (1.00% + 2d) ± (0.60% ÷ 2d)
4M0 ± (1.00% + 4d) ± (0.60% + 4d)
40M0 ± (7.00% + 5d) ± (7.00% + 5d)
Resolution: 0.01 Ω in the 4000 range
Open Circuit Voltage: 3.3V
Open Circuit Low Voltage: 0.6V
Input Protection: 600V rms
CONTINUITY CHECK
Continuity Threshold: Approximately 50Ω
Continuity Indicator: 2KHz Tone Buzzer
Input Protection: 600V rms
DIODE TEST
Test Current: 1.1 mA (Typical)
Open Circuit Voltage: 3.3V DC (max)
Input Protection: 600V rms
Form Number TM61633-5 Rev 5 Dec 2006