IDEAL INDUSTRIES Meter Thermometer User Manual


 
Page 4
Measurement Characteristics
All at 23°C ±5°C, < 80% Rh. Specifications are expressed as ±([% of reading] + [number of digits]).
Multiply accuracy Digits by l0 in 40,000 count Mode.
DC VOLTAGE:
DCV 61-633 61-635
40mV ± (0.20%+8d) ± (0.06% + 8d)
400mV ± (0.20% + 2d) ± (0.06% + 2d)
4V, 40V, 400V, 1000V ± (0.20% + 2d) ± (0.06% + 1d)
AC VOLTAGE:
ACV 61-633 61-635
400mV
40Hz~ 100Hz ± (1.20% + 5d) ± (0.70% + 5d)
100Hz ~ 1KHz ± (2.00% + 5d) ± (1.00% + 5d)
4V
40Hz ~ 100Hz ± (1.00% + 5d) ± (0.70% + 5d)
100Hz ~ 1KHz ± (2.00% + 5d) ± (1.00% + 5d)
1KHz ~ 10KHz ± (3.00% + 6d) ± (2.00% + 6d)
10KHz ~ 20KHz N.A ± (3.00% + 7d)
20KHz ~ 50KHz N.A ± (5.00% + 8d)
50KHz ~ 100KHz N.A ± (10.00% + 10d)
40V
40Hz ~ 100Hz ± (1.20% + 5d) ± (0.70% + 5d)
100Hz ~ 1KHz ± (2.00% + 5d) ± (1.00% + 5d)
1KHz ~ 10KHz ± (3.00% + 6d) ± (2.00% + 6d)
10KHz ~ 20KHz N.A ± (3.00% + 7d)
20KHz ~ 50KHz N.A ± (5.00% + 8d)
50KHz ~ 100KHz N.A ± (10.00% + l0d)
400V
40Hz ~ 100Hz ± (1.20% + 5d) ± (0.70% + 5d)
100Hz ~ 1KHz ± (2.00% + 5d) ± (1.00% + 5d)
1KHz ~ 10KHz ±(3.00% + 6d) ± (2.00% + 6d)
10KHz ~ 20KHz N.A ± (3.00% + 7d)
20KHz ~ 50KHz N.A ± (5.00% + 8d)
750V
40Hz ~ 100Hz ±(1.20% + 5d) ± (0.70% + 5d)
100Hz ~ 1KHz ± (2.00% + 5d) ± (1,00% + 6d)
Bandwidth 40Hz ~ 10KHz 40Hz ~ 100 KHz
dBm (typical): -l5dBm to +55 dBm (0 dBm = lmW into 600Ω).
dBv (typical): -80 dBv to + 50 dBv (0 dBv = 1 Vrms).
Note: (ACV only)
Add additional 40Digits for reading under 30% of range.
Specifications exclude under 20% of range for 20KHz ~ 100KHz.
Resolution: 1µV in the 400mV range
Input Impedance: 10MΩ, <l00pF
Overload Protection: 1000V dc, 750V rms
AC Conversion Type: AC Coupled True RMS responding
AC + DC Volts: Same as AC(RMS) +1.00% + 8d
Crest Factor: +1.5% addition error for C.F. 1.4 to 3
+3.0% addition error for C.F. from 3 to 4
Form Number TM61633-5 Rev 5 Dec 2006