9
Humidity Effect
Humidity has no effect on the performance of the MAXO
2
analyzer other
than diluting the gas, as long as there is no condensation. Depending on
the humidity, the gas may be diluted by as much as 4%, which
proportionally reduces the oxygen concentration from the dry
concentration. Environments where condensation may occur are to be
avoided since condensate may obstruct passage of gas to the sensing
surface, resulting in erroneous readings and slower response time. For
this reason avoid usage in environments greater than 95% relative
humidity.
3. OPERATION INSTRUCTIONS
To check the oxygen concentration of a sample gas: (after the unit has
been calibrated)
1) Connect the sample hose to the barbed fitting on the bottom of the
unit.
2) Attach the other end of the sample hose to the sample gas source
and initiate flow of the sample to the unit at a rate of 1-5 liters per
minute (2 liters per minute is recommended).
3) Using the ON/OFF key, make sure the unit is in the power on mode.
4) Allow the oxygen reading to stabilize. This will normally take about
30 seconds or more.
4. SENSOR REMOVAL AND REPLACEMENT
The OM-25A is shipped with a new MAX-250 oxygen sensor installed.
Although the sensor has a very long expected life, eventually the sensor
will require replacement. Removing or installing a sensor, when
necessary, is a very simple procedure. To remove and install a new
sensor:
1) With the front of the analyzer facing downward and the bottom facing
away from you, press down with both thumbs on the center of the
battery compartment cover and slide the cover off.
2) Remove the white barb fitting by turning the barb counter-clockwise.
A
7
/16” wrench may be needed to aid in the removal.
R213M15K.qxd 6/26/2006 8:59 AM Page 11