Health O Meter 498KL Scale User Manual


 
3
CAUTION AND WARNING
To prevent injury and damage to your scale, please follow these instructions very carefully:
Do not transport the scale while platform is loaded.
For accurate weighing the scale must be placed on a flat, stable surface.
For accurate weighing, verify proper operation according to the procedure described in this manual before each use.
Do not use in the presence of flammable materials.
Operating at other voltages and frequencies than specified could damage the equipment.
If the “LOW BAT” indicator activates, for accurate weighing replace the batteries or connect the scale to an AC power source
as soon as possible.
Only manufacturer’s qualified personnel shall service the scale.
This equipment has been tested and found to comply with the EMC limits for the Medical Device Directive 93/42/EEC (EN
55011 Class A and EN 60601-1-2). These limits are designed to provide reasonable protection against harmful interference in
a typical medical installation. The equipment generates, uses and may radiate radio frequency energy and, if not installed and
used in accordance with these instructions, may cause harmful interference to other devices in the vicinity. However, there is
no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference with
other devices, which can be determined by turning the equipment off and on, the user is encouraged to try to eliminate the
interference by one or more of the following measures:
Reorient or relocate the receiving device.
Increase the separation between the equipment.
Connect the equipment into an outlet on a circuit different from that to which the other device(s) is connected.
Consult the manufacturer or field service technician for help.
DEFINITION OF SYMBOLS
498KL SCALE SPECIFICATIONS
Capacity and Resolution: 500lb x 0.2lb / 220kg x 0.1kg
Power Requirements: 6-AA cell 1.5V batteries (INCLUDED) or optional Adapter Model RHD090020 / ADPT31 (USA -
CSA only) 120v AC – 9V DC 60 Hz or with Adaptor Model UE 15WCP-090050SPA / ADPT30
(IEC only) 120-240V AC – 9V DC 50-60Hz. OPTIONAL ADAPTERS NOT INCLUDED.
Environmental: Operating temperatures: 50°F to 104°F / 10°C to 50°C
Atmospheric Pressure: 500hPa to 1060hPa
Humidity: 10% to 95% RH
Platform Dimensions: 17 ¾” x 14” x 2 ¼” / 43cm x 35.6 cm x 5.8 cm
Remote Indicator Dimensions: 8” x 4 ½” x 2” / 20.4cm x 11.7cm x 4.9cm with 1” / 2.54cm LCD
STORAGE SPECIFICATIONS
Storage Temperatures: 32°F to 122°F / 0°C to 50°C
Humidity: 10% to 95% RH
Atmospheric Pressure: 500hPa – 1060hPa
498KL SCALE CONNECTIVITY INFORMATION:
This scale is designed to connect to other devices via a USB port, to allow acquisition of weight, height and BMI from the
scale, to help close a source of potential medical errors caused by incorrectly copying and then recording a weight taken.
This scale is EMR-EHR compatible. For more information regarding how to connect this scale to your computer, a
monitor, printer or other device please contact Health o meter® Professional Technical Support at 800-638-3722.
*WARNING:
For use with AC Adaptor Model RHD090020 (USA - CSA only) or with Adaptor
Model UE 15WCP-090050SPA/ADPT30 (IEC only). Complies with C22.2
No. 601.1-M90 and UL Std No. 60601-1. Type BF applied parts. Internally powered.
Equipment is not suitable for use in the presence of flammable anesthetic mixture
with air
n
or with oxygen or nitrous oxide.
AC to DC power supply
(9V DC, 200mA)
DC Connector Polarity
Temperature range
(50°F to 104°F / 10°C to 50°C)
Humidity range
(10% - 95% RH)
Atmospheric pressure range
(500hPa – 1060hPa)
Attention, Consult Accompanying Documents
Type BF Applied Parts
AC to DC power supply