Graham Field US0458 Personal Lift User Manual


 
US0208-INS-LAB-RevD11 • Patriot Homecare Bed 7
Install motor slide caps
7. Position the caps as shown at right so that
the grooves in the slide cap are aligned with
the grooves in the motor.
8. Apply straight downward pressure to each
end of the cap while sliding into the closed
position.
9. Repeat step 8 for the other cap.
10. Ensure that all four corners of each slide cap
are engaged.
WARNING: Ensure that the motor is assem-
bled on bed as instructed above before use.
Motor assembly troubleshooting
Problems associated with improper motor installation are listed below, with appropriate x.
Problem Probable cause Fix
Head operates with knee control button and
knee operates with head control button
Motor is installed backwards Remove motor and reverse it (see previ-
ous page)
Knee lifting linkages bind against grid
deck, preventing knee deck from elevating.
Motor is installed backwards, enabling
knee to elevate too high and bind linkage
Remove motor and reverse it (see previ-
ous page)
motor
slide cap
Install motor slide caps
pressure