Fluke 117 Thermometer User Manual


 
114, 115, 116, and 117
Calibration Information
4
Accuracy Specifications
Function Range Resolution
Accuracy
± ([% of Reading] + [Counts])
Model
[1]
DC millivolts 600.0 mV 0.1 mV 0.5 % + 2 114, 115, 116, 117
DC Volts
6.000 V
60.00 V
600.0 V
0.001 V
0.01 V
0.1 V
0.5 % + 2 114, 115, 116, 117
DC, 45 to 500 Hz 500 Hz to 1 kHz
Auto-V LoZ
[2]
True-rms 600.0 V 0.1 V 2.0 % + 3 4.0 % + 3 114, 116, 117
45 to 500 Hz 500 Hz to 1 kHz
AC millivolts
[2]
True-rms 600.0 mV 0.1 mV 1.0 % + 3 2.0 % + 3 114, 115, 116, 117
AC Volts
[2]
True-rms
6.000 V
60.00 V
600.0 V
0.001 V
0.01 V
0.1 V
1.0 % + 3 2.0 % + 3 114, 115, 116, 117
Continuity 600 Ω 1 Ω
Beeper on <20 Ω, off >250 Ω; detects opens
or shorts of 500 μs or longer.
114, 115, 116, 117
Ohms
600.0 Ω
6.000 kΩ
60.00 kΩ
600.0 kΩ
6.000 MΩ
40.00 MΩ
0.1 Ω
0.001 kΩ
0.01 kΩ
0.1 kΩ
0.001 MΩ
0.01 MΩ
0.9 % + 2
0.9 % + 1
0.9 % + 1
0.9 % + 1
0.9 % + 1
1.5 % + 2
114, 115, 116, 117
Diode test 2.000 V 0.001 V 0.9 % + 2 115, 116, 117
Capacitance 1000 nF
10.00 μF
100.0 μF
9999 μF
1 nF
0.01 μF
0.1 μF
1 μF
1.9 % + 2
1.9 % + 2
1.9 % + 2
100 μF - 1000 μF: 1.9 % + 2
>1000 μF: 5 % + 20
115, 116, 117
Lo-Z Capacitance (Power-
up option)
1 nF to 500 μF
20% + 2
(10% +2 typical)
115, 116, 117
AC Amps True-rms
[2]
(45 Hz to 500 Hz)
6.000 A
10.00 A
[4]
20 A overload for 30
seconds maximum, 10
minutes rest minimum.
0.001 A
0.01 A
1.5 % + 3 115, 117
DC Amps
6.000 A
10.00 A
[4]
20 A overload for 30
seconds maximum, 10
minutes rest minimum.
0.001 A
0.01 A
1.0 % + 3 115, 117
Temperature (Type K
thermocouple)
-40 °C to 400 °C
-40 °F to 752 °F
0.1 °C
0.2°F
1.0 % + 10
[5]
1.0 % + 18
[5]
116
AC μAmps True-rms
[2]
(45 Hz to 1 kHz)
600.0 μA 0.1 μA
1.5 % + 3 (2.5 % + 3 > 500 Hz) 116
DC μAmps 600.0 μA 0.1 μA 1.0 % + 2 116