Care and Maintenance
• Accuracy Limits: Comparison between the reference instrument(s) and the
instrument being verified should be within ±0.4 ºF (0.2 ºC) for acceptable limits.
If not, repeat the process. In the event they still differ by more than the acceptable
limits, call Exergen for repair or replacement of the failed instrument.
Verifier Specifications:
Power Source 9-volt battery, or 9-volt power supply.
Battery Life Approximately 1 hour continuous use.
Low Voltage Indicator Red LED shuts off when battery voltage drops below
~5 volts.
Temperature Range 97-104 ºF (36-40 ºC)
Cleaning Wipe down with alcohol or any hospital approved
disinfectant. Do not immerse.
Repair
If repair is required:
• Contact Exergen at (617) 923-9900 for a Return Materials
Authorization (RMA) Number.
• Mark the RMA number on the outside of your package and packing
slips.
• Include a description of the fault if possible.
• Send the instrument freight/postage prepaid to:
Exergen Corporation
400 Pleasant Street
Watertown, MA 02472
• The instrument will be returned freight/postage prepaid.