Drive Medical Design 18605 Oxygen Equipment User Manual


 
12
specifications / classification
APPROVALS
18605 Series . . . . . . . . . . . . . . . . . . . . . . . . . . . .IEC 601-1; CAN/CSA-C22.2 No. 601.1-M90; UL 60601-1,
EN 60601-1-2
18605 meets RTCA/D-160D . . . . . . . . . . . . . . . . . . . . . . . . .DO-160D - section 21 Category M
EQUIPMENT CLASSIFICATIONS
With respect to protection from electric shock . . . . . . . . . . . .Class I and internally powered
Degree of protection against electric shock . . . . . . . . . . . . . .Type BF Applied Parts
Degree of protection against ingress of liquids . . . . . . . . . . . .IPX2 and ordinary power supply
Equipment not suitable for use in the presence of a flammable anaesthetic mixture with air or
nitrous oxide.
ISO Classification
18605 Series only - Electrically powered medical suction equipment for field and transport use
according to ISO 10079-1 : 1991 High Flow/High Vacuum
18605Series - Electrically powered medical suction equipment for non-transport use according
to ISO 10079-1 : 1991
Drive Medical will make available on request circuit diagrams, parts lists, etc.