Drive Medical Design 18605 Oxygen Equipment User Manual


 
10
maintenance & troubleshooting
MAINTENANCE
Inspect suction tubing and collection bole for leaks, cracks, etc. before each use.
DANGER
Electric shock hazard. Do not aempt to open or remove cabinet, there are no user-serviceable
internal components. If service is required, return unit to a qualified Drive Medical provider or an
authorized service center. Opening or tampering with the unit will void warranty.
Filter Changing:
1. Change bacteria filter if overflow occurs or every two months, whichever comes first.
2. Remove filter by disconnecting it from suction unit and lid assembly.
3. Replace with a clean Drive bacteria filter (non-sterile) (10/pack) and remount to suction unit
and lid. Additional filters may be purchased from your authorized Drive Medical provider.
NOTE
Do not substitute any other material for this bacteria filter. Substitution may lead to contamina-
tion or poor performance; use only Drive filters.
TROUBLESHOOTING
NOTE: Your Drive Suction Unit contains no user-serviceable parts. If you believe your unit is not
working properly, BEFORE YOU RETURN IT TO THE HOME MEDICAL EQUIPMENT PROVIDER
WHERE YOU PURCHASED IT OR TO Drive MEDICAL, please take a few moments to check for
these possible causes:
PROBLEM ACTION
Unit does not turn on. 1. Check power sources and connections.
(Green external power indicator 2. Ensure wall outlet is live by plugging in a lamp.
should be illuminated when 3. If running from an internal baery, ensure that your
power is applied.) unit has a baery installed.
4. If baery is installed, check that it is fully charged.
Pump runs, 1. Check that all tubing is connected properly.
but no vacuum. 2. Check tubing connections for breaks or leaks.
3. Ensure that flow shut-off is not activated.
4. Check for leaks or cracks in bole assembly.
Low vacuum. 1. Use vacuum adjustment knob to increase vacuum level.
2. Check system for leaks.
3. Push vacuum adjustment knob and then release.
Baery will not hold a charge. 1. Ensure that unit is equipped with an internal baery
(Charge indicator should be by contacting your Drive Medical equipment provider.
illuminated if baery is connected 2. Verify that charge light turns on.
during charge mode.) 3. Check electrical connections during charging.
4. Ensure wall outlet is live by plugging in a lamp.