Drive Medical Design 18605 Oxygen Equipment User Manual


 
9
cleaning &maintenance
Collection Bole:
1. Shut off unit using power switch and allow vacuum to drop. Disconnect power source from the
DC input receptacle on the unit.
2. With the collection bole still in the holder, remove the lid. The bole can now be taken out of
the holder to be emptied.
NOTE
Collection bole should be emptied and cleaned aer each use. The 800ml disposable collection
bole and lid are meant for single-patient use
3. Collection boles and lids should be thoroughly cleaned aer each use by one of the following
methods:
All Boles:
. Wash in a hot water/dishwashing detergent solution and rinse with clean, hot tap water. Then
wash in one part vinegar to three parts hot water solution. Rinse with hot tap water and air dry.
Autoclave for 3 to 5 minute cycle at 270
o
F (132
o
C).
In lieu of autoclave sterilization, the above parts can be soaked in an activated dialdehyde
solution.
Suction Unit:
1. With the power switch in the “OFF” position, disconnect the Drive Suction Unit from all exter-
nal power sources.
2. Wipe the housing with a clean cloth and any commercial (bacterial-germicidal) disinfectant.
CAUTION
Do not submerge in water as this will result in damage to the vacuum pump.
Tubing:
1. Disconnect the tubing from the unit.
2. Tubing should be rinsed thoroughly aer every use by running hot tap water through it fol-
lowed by a solution of one part vinegar to three
parts hot water.
3. Rinse with hot tap water and air dry.
4. Keep the outer surface of the tubing clean by wiping with a clean, damp cloth.
Carrying Case:
1. Wipe the case using a clean cloth dampened with detergent and/or disinfectant.