Delphi DL-00183 Oxygen Equipment User Manual


 
21
DL-00096 Rev. 7
Appendix 1: Technical Description
Size: 7.4”Wx4.6”Dx11.6”H(188mmWx117mmDx295mmH)
Unit weight: Lessthan10lbs(withoutcarryingcaseorcart)
Power requirements: 100-240VAC(±10%),50-60Hz;batterypower18VDC,6.7amps;DCadap-
tor,18VDC,6.7amps(Note: See Approved Accessories Guide for model and part number of AC
power supply.)
Purity: 90%±3%(22°C±3°C,14.29psia±0.04or739mmHgat40%±15%relativehumidity)
at all flow rates
Setting: Useradjustable1.0to5.0(with0.5increments)(min.1.0;max.5.0)
Setting indicator: LCDdisplay
Maximum oxygen discharge pressure:17psiatsettingof5.0
Humidity range: 5%to93%±2%non-condensing
Operating altitude: 0’to8,000’relativetosealevel(0kmto2.4km)
Sound pressure level (measured 1 m from edge of chassis):Lessthan50dB(A)average
Type of protection (electrical): ClassI
Degree of protection (electrical): TypeBF
Degree of protection (water):IPX2
Degree of safety (flammable anesthetic mixture): Notsuitableforuseinthepresenceofa
flammableanestheticmixture
Operation: Continuousoperationattemperaturesof95°F(35°C)downto41°F(5°C),andshort-time
operationfor25minutesattemperaturesupto104°F(40°C).
Alarm sound pressure range: 65to85dB(A)
Alarm system delays: Lessthan10secondsafterdetection(lowoxygenalarmsifoxygenisless
than82%volumefractionatspecifiedenvironmentalconditions)
Oxygen concentrator status indicator:Highpriorityalarmthatindicateswhenoxygenconcentra-
tiondropsbelow82%
Maximum Temperature:Underextremeenvironmentalandoperatingconditions,specificmaxi-
mumsurfacetemperaturesarelistedbelow.Eachofthesespecificsurfacesalsolistsanexpected
user contact limit as follows:
TheKeypanelcanreachupto140°F(60°C).Thissurfaceisintendedtobetouchedforlessthanone
minuteatatime.Theclearplasticwindowonthecarryingcasecanreachupto118.4°F(48°C).This
surfaceisintendedtobetouchedforlessthantenminutesatatime.Theremainderofthecarrying
casecanreachupto109.4°F(43°C).Thissurfacehasnotimelimitationonusercontact.
DL-00096 rev 7.indd 21 6/1/2009 4:48:34 PM