Delphi DL-00183 Oxygen Equipment User Manual


 
16
DL-00096 Rev. 7
IfthePortableOxygenConcentratordetectsanalarmcondition,itwillindi-
catethealarmvisuallyandaudiblywithin10seconds.Therearefourlevels
ofalarms:criticalhighpriority,highpriority,mediumpriority,andlowpriority.
Eachisindicateddifferentlybythebacklitdisplay;green,yellow,andred
LEDs;andbuzzer,asindicatedbelow.Ineachcase,thealarmmessageand
powerstatuswilloverridethecurrentdisplay.
Note: All alarm conditions and parameters are factory preset; conditions and parameters cannot be
changed or adjusted by the user.
Alarm Status
Criticalhigh
priority
High priority
Mediumpriority
Lowpriority
Audible Tone
Ten beeps per burst, burst
repeatsevery3seconds
Ten beeps per burst, burst
repeatsevery3seconds
Three beeps per burst,
burstrepeatsevery8
seconds
Three beeps per burst,
burstrepeatsevery
10minutes
Visual Indicator
SolidredLED
(anddeviceshuts
offautomatically)
FlashingredLED
FlashingyellowLED
SolidyellowLED
Note: If two alarm conditions exist at the same time, the highest priority alarm is indicated. If two or
more alarm conditions of equal priority exist at the same time, the most recent one will be displayed.
Note: The most recent alarms indicated by the device are logged for reference by service personnel.
This log is maintained even if the device is powered down or if power is lost for any other reason.
Note: If the mute key is pressed prior to an alarm condition (for example, to mute the device in a
movie theater), critical high priority and high priority alarms will override the mute function; medium
and low priority alarms will be muted for eight hours from the time the key was pressed. Press the
mute key off to display the last highest priority alarm. Press the mute key on again to reset the eight-
hour timer.
ALARMS
WhenthePortableOxygenConcentratorsoundsanalarm,acorresponding
messagewillbedisplayedonthescreen.Takeappropriateactionasdirected
inthechartsbelow.
Critical High Priority Alarms
Note: These alarms will disable the Portable Oxygen Concentrator immediately.
Alarm Message
ChargeBattery
XX:Service!*
*Value:01-20
Description
Batteryneedscharging.
Servicerequired.
Action
Recharge the battery
pack by plugging in to
powersupply.Ensureall
connections are made
securely.
ContactCustomer
Service.
Alarm Indicators
Mute Time
20minutes
20minutes
8hours
24hours
DL-00096 rev 7.indd 16 6/1/2009 4:48:34 PM