Delphi DL-00183 Oxygen Equipment User Manual


 
18
DL-00096 Rev. 7
Alarm Message
ReplaceC-filtr
ReplaceP-filtr
WashI-filtr
Replace batt
XX:Service*
*Value:71-99
Description
Compressorfilterneeds
replacement.
Patientfilterneeds
replacement.
Inputfilterneedstobe
cleaned.
Mainbatteryhasexceeded
300rechargecycles.
Servicerequired.
Action
Replacefilter.Pressand
hold mute key for more
than4secondstoreset
thehourcounter.
Replacefilter.Pressand
hold mute key for more
than4secondstoreset
thehourcounter.
Washinputfilter.Press
and hold mute key for
morethan4secondsto
resetthehourcounter.
ContactCustomer
Serviceforreplacement.
Note: This alarm will be
automatically cleared once
a new battery is inserted.
Battery will continue to
function after 300 recharge
cycles, but for a shorter time
than a new battery.
ContactCustomer
Service.
Low Priority Alarms
Other Messages
Message
Powerremoved
SWVerN.NNN
Self test
Shutting off
Externalpower
Description
Externalpowerhasbeen
disconnected; unit is now
runningonbattery.
Displayofsoftwareversion
during startup and calibra-
tion.Thisincludestestof
beeperandLEDs.
Indicatesunitisgoing
throughitstestatstartup.
Displayedforapproximately
10seconds.
Displayedwhileunitgoes
throughitspower-down
sequence.
Externalpoweris
connected.
Action
Noactionisrequired.
Noactionisrequired.
Noactionisrequired.
Noactionisrequired.
Noactionisrequired.
DL-00096 rev 7.indd 18 6/1/2009 4:48:34 PM