Bayer HealthCare Elite XL Blood Glucose Meter User Manual


 
5
section
27
Glucometer Elite
®
XL
section
27
Test Sample: Whole blood
Test Result: Plasma/serum glucose
Reaction Chamber
Sample Volume: 2 L
Measuring Range: 20–600 mg/dL
(1.1–33.3 mmol/L)
Measuring Time: 30 Seconds
Date: Year 2000 compliant
Calibration Curve: Selected by use of the
Code Strip
Memory Feature: Stores last 120 test results
Temperature
Compensation: Automatic compensation
with built-in thermistor
Battery Type: 3-volt lithium batteries
(DL or CR2032) – 2
Battery Life: Approximately 1000 Tests
(1 year of average use)
Operating:
Temperature Range: 50°–104° F (10°–40° C)
Humidity: 20–80% RH
Dimensions: 97.8 x 56 x 14.5 mm
(3
7
8 x 2
3
16 x
9
16 in.)
Weight: Approximately 60g
(2.1 ounces)
Alarm: “Beeps” whenever a Check
Strip, Code Strip or Test
Strip is inserted into the
Meter, when the Test Strip
is filled with blood or
Control Solution, and
when a test result appears
in the display.
Specifications
I4