Advanced Adjustments 20
Menu exit
1 Turn the Brightness knob to enter the Exit
menu. The Save Changes dialog appears.
2 Turn the Brightness knob to toggle between
Yes and No. Select Yes to save the changes,
select No to exit without saving the changes.
3 Turn the Contrast knob to exit.
Note:
A number of settings and adjustments are saved
directly after changing them. They are: ALC on/
off, Backporch clamp on/off, Remote timeout,
Luminance uniformity, Orbiter, Power save,
Address, Rotation, and all Southern Hemisphere
adjustments.
Mode selection
This is the indication of the current scanning
mode. If the scanning mode system is set to
Manual (see further), you can select another
scanning mode by turning the Brightness knob.
Turn the Brightness knob to toggle the Automatic
scanning system between Off and On.
When switched to Off, the scanning mode
selection is manual.
Turn the Brightness knob to toggle the automatic
Power Save system between Off and On.
Turn the Brightness knob to set the Remote
time-out. This is the time, expressed in seconds,
before the display interrupts the communication
with the computer if the computer does not
respond anymore.
Turn the Brightness knob to set the display’s
address. This is necessary if several displays are
controlled by the same computer (by means of
MediCal). In that case, every display must have a
different address.