A&D UA-767 30 Blood Pressure Monitor User Manual


 
Maintenance
Do not open the device. It uses delicate electrical components and an
intricate air unit that could be damaged. If you cannot fix the problem
using the troubleshooting instructions, request service from your dealer
or from the A&D service group. The A&D service group will provide
technical information, spare parts and units to authorized dealers.
The device was designed and manufactured for a long service life.
However it is generally recommended to have the monitor inspected
every 2 years, to ensure proper functioning and accuracy. Please contact
either your authorized dealer or A&D for maintenance.
Technical Data
Type UA-767 30
Measurement method Oscillometric measurement
Measurement range Pressure: 20 - 280 mmHg
Pulse: 40 - 200 beats / minute
Measurement accuracy Pressure: ±3 mmHg or 2%, whichever is greater
Pulse: ±5%
Power supply 4 x 1.5V batteries (R6P, LR6 or AA)
Upper arm circumference 22 - 32 cm
Classification Type BF
Clinical test According to ANSI / AAMI SP-10 1987
EMC IEC 60601-1-2: 2001
Memory Last 30 measurements
Operating condition +10°C to +40°C / 30%RH to 85 %RH
Storage condition -10°C to +60°C / 30%RH to 85 %RH
Dimensions Approx. 147 [W] x 64 [H] x 110 [D] mm
Weight Approx. 300 g, excepting batteries
Note: Specifications are subject to change without prior notice.
English 13