25
11-2. Function list
Item
Function number
Description
f 1- 0
Auto power-off disabled
Auto power-off
function
f 1- 1
Auto power-off enabled
Automatically
power off
f 2- 0
Normal (1/3,000 class)
f 2- 1
High (1/6,000~1/7,500 class)
Display
resolution
f 2- 2
Higher (1/12,000~1/15,000 class)
Legal for trade is
Normal only.
f 3- 0
kg
f 3- 1
g
f 3- 2
lb
f 3- 3
oz
Weighing unit
when powered
on
f 3- 4
lb-oz
Factory setting
may be different
depending on the
area to be
shipped.
f 4- 0
2400 bps
f 4- 1
4800 bps
Serial interface
Baud rate
f 4- 2
9600 bps
Optional RS-
232C/422/485
f 5- 0
7 bit / Even parity
f 5- 1
7 bit / Odd parity
Serial interface
Data bit / Parity
f 5- 2
8 bit / Non parity
Optional RS-
232C/422/485
f 6- 0
Stream mode
f 6- 1
Command mode
f 6- 2
Print key mode
f 6- 3
Auto-print mode + data
Serial interface
Data output
mode
f 6- 4
Auto-print mode +/- data
Optional RS-
232C/422/485.
The UFC format
is applicable to
f6-2 to 4.
f 7- 0
Upper & Lower limits weight setting
mode
f 7- 1
Target weight setting mode with
HI/LO limit weight
Comparison
mode
f 7- 2
Target weight setting mode with
HI/LO limit % of the target weight
Way to set
parameters.
f 8- 0
Comparator disabled
f 8- 1
Compares all data
f 8- 2
Compares all stable data
f 8- 3
Compares data > +4d or < -4d
f 8- 4
Compares stable data > +4d or < -4d
f 8- 5
Compares data > +4d
Comparator
operating
condition
f 8- 6
Compares stable data > +4d
Conditions to
compare.
d = minimum
display division
f 9- 0
Buzzer does not sound.
f 9- 1
Buzzer sounds at LO
f 9- 2
Buzzer sounds at OK
f 9- 3
Buzzer sounds at LO and OK
f 9- 4
Buzzer sounds at HI
f 9- 5
Buzzer sounds at LO and HI
f 9- 6
Buzzer sounds at OK and HI
Comparator
buzzer
f 9- 7
Buzzer sounds at LO, OK and HI
Buzzer sounds
according to
the comparator
results
f10- 0
f10- 1
f10- 2
f10- 3
Response /
Filtering
f10- 4
Fast / Weak (sensitive)
Slow / Strong (stable)
Software
filtering
Factory setting