A&D Ambulatory Blood Pressure Monitor Blood Pressure Monitor User Manual


 
Page 32
Error Code Maintenance
Error codes
Caution The error code updates without announcement.
Error
code
Meaning
Status Operation and Treatment
E00
E03
E04
E05
E06
E07
E08
E10
E20
E21
E22
E23
No clock
parameter
Pressure zero
error
Low battery
Inflation error
Above 320mmHg
Controlled stop
using STOP key
Pulsation can not
be measured
Pulsations can not
be detected
because the
patient may have
moved.
Pulse rate < 30
200 < Pulse rate
DIA < 40
160 < DIA
SYS < 60
280 < SYS
DSD < 10
150 < DSD
Enter clock parameters. Refer to
"Setup of display and clock"
Release the air from the cuff
completely.
Replace with new batteries.
Restart the auto mode if you use it.
Wrap the cuff and connect to main
unit exactly. If you can not clear the
error, there may be an air leak and
repair is necessary
Do not move and try to relax during
the measurement. If you can not
clear the error, the product will
requier repair.
Do not press the STOP key if you do
not need to use it.
Do not move and try to relax during
the measurement. The error occurs
when measurable pulsations are not
received due to thick cloth or quick
motion.
Do not move and try to relax during
the measurement.
Measure the blood pressure by other
methods.
All parameters are lost.
Reset status.
An error code is displayed
without cuff inflation.
Measurement is stopped.
An error code is displayed.
Auto mode is quit.
Inflation pressure does not
reach the target pressure.
An error code is displayed.
Air is exhausted. An error
code is displayed.
Measurable pulsation is
searched to 20mmHg in
constant exhaust. An error
code is displayed.
In the measurement, Quick
exhaust is executed. An
error code is displayed.
An error code is displayed.
DIA
SYS
DSD
: Diastolic Blood Pressure
: Systolic Blood Pressure
: The Difference between Systolic Blood Pressure
: and Diastolic Blood Pressure.