Section 5 - Care and Maintenance
5.4 Specifications
Admissible ambient conditions
Use only in closed rooms
Temperature range: 5 ºC to 40 ºC
Atmospheric humidity: 80% rh @ to 30 ºC
Warm-up time: At least 60 minutes after connecting the instrument to the power supply;
when switched on from standby-mode, the instrument is ready for
operation immediately.
Voltage fluctuations: 15% +10%
Installation category: II
Pollution degree: 2
Power load: Max. 450 W during drying process
Current consumption: 4 A or 2 A, according to the heating element
Power supply voltage: 100 V  120 V or 200 V  240 V, 50/60 Hz
(the voltage is given by the heating element)
Power line fuse: 1 piece, 5 x 20 mm, T6,3 H 250 V
MODEL MB35
Capacity 35g
Readability 0.005g, 0.05%
Temp. Settings 50°C to 160°C (5° increments)
Drying Programs Standard
Switch-off Criteria Timed, Auto
Heat Source Halogen
Calibration External calibration mass-20g
Dimensions (DxWxH) (in/cm) 14 x 7.5 x 6/ 35.5 x 19 x 15.2
Pan Size 90 mm. diameter
Weight (lb/kg) 9.8 / 4.5
Shipping Weight (lb/kg) 14.1 / 6.4