Section 3 - Analyzer Setup
3.3 Temperature Calibration
ANALYZER SETUP
TEMP CAL
REMOVE PAN SUPPORT
ABORT CALIBRATION
WEIGHT CAL
TEMP CAL
TIME-DATE
RS-232
PRINT
EXIT
NOTE: Optional Temperature Calibration Kit is required.
1.
TARE
2.
ANALYZER SETUP
WEIGHT CAL
TEMP CAL
TIME-DATE
RS-232
PRINT
EXIT
TEMP CAL
INSERT CAL FIXTURE
CLOSE COVER
ABORT CALIBRATION
TO START
TEMP CAL
CALIBRATION READY
PRESS TARE/ACCEPT
ABORT CALIBRATION
TEMP CAL 100C
CURRENT TEMP: --- C
TM TO CAL PNT: 15:00 MIN
ADJ CAL READING: 100 C
ACCEPT NEW CAL
ABORT CALIBRATION
TARE
ADJUST TO THERMOMETER READING
TEMP CAL 100C
CURRENT TEMP: 100 C
TM TO CAL PNT: 10:00 MIN
ADJ CAL READING: 100 C
ACCEPT NEW CAL
ABORT CALIBRATION
TEMP CAL 100C
CURRENT TEMP: 100 C
TM TO CAL PNT: 10:00 MIN
ADJ CAL READING: 101 C
ACCEPT NEW CAL
ABORT CALIBRATION
TEMP CAL 100C
CURRENT TEMP: 100 C
TM TO CAL PNT: 09:23 MIN
ADJ CAL READING: 101 C
ACCEPT NEW CAL
ABORT CALIBRATION
PREPARATION
TARE
3.
START CALIBRATION
100C
15 minute time lapse
4.
AFTER 15 MINUTE TIME PERIOD
x1
or