14
Assembly
Supply unit for lamp housing 106Z
with 50W mercury-vapour burner
1. On/off switch
2. Running-time meter
3. Selector for L1/L2 burners
4. Fuse
5. Selectorfor50Hz/60Hz
6. Connector for lamp
7. Power cable
The supply unit is intended for power supplies above 220V.
To run it on lower voltages a step-down transformer is required
(e.g. 230V to 110V).
Observe the safety regulations on page 5.
▶ Connect the cable from the lamp housing to the supply unit
▶ Engage “L1” or “L2” in accordance with the inscription
on the burner socket (see page 12)
▶ Engage50Hzor60Hz
▶ Connect the power cable to the supply unit and to the grid
▶ Make a note of the running-time meter reading.
Supply unit for lamp housing 106Z
with 100W mercury-vapour burner
1. On/off switch
2. Pilot lamp LAMP
3. Pilot lamp TEMP
4. Running-time meter
5. Pilot lamp SAFETY
Observe the safety regulations on page 5
▶ Connect the cable from the lamp housing to the supply unit
▶ Secure the lamp cable.
▶ Connect the power cable to the supply unit and to the grid
▶ Make a note of the running-time meter reading.
2 1 6 7
543
Input type
220V/240V±10%
220V~50/60Hz
Hg50WL1/L2
Osram no. 0-958
Max. power 320VA
Type ebq 100dc
Input voltage 90V to 250V AC, max 265VA
Inputfrequency 48Hzto63Hz
Protection class IP20
Series no. G 34240090
Fuses
The fuse holder is secured in the holder with spring clips
on both sides.
▶ Insert a screwdriver behind the spring clips and extract
the fuse holder.
▶ Insert the fuses (2×T3,15AH250VF1/F2).
▶ Refit the fuse holder
1
3
2
4 5