Programming Reference
SCPI Instrument Command List
Command :OUTP[1]
Long
:OUTPut[1][:NORMal][:STATe]
Form Set & Query
Parameter
ON | OFF | 1 | 0
*RST value
OFF
Description Use this command to switch the normal OUTPUT on or off.
Example To switch on the output:
:OUTP ON
Command :OUTP[1]:IMP
Long
:OUTPut[1]:IMPedance[:INTernal]
Form Set & Query
Parameter Numeric
Parameter Suffix
OHM
with engineering prefixes, e.g.:
MOHM
is Megaohms.
*RST value 50
Specified Limits 50
or 1 k
Description Use this command to program the source impedance of the OUTPUT
connector. Note that only two settings are available. If you try to program
any other value, it will be rounded to one of the specified values.
Example To program the source impedance:
Command :OUTP[1]:IMP:EXT
Long
:OUTPut[1]:IMPedance:EXTernal
Form Set & Query
Parameter Numeric
Parameter Suffix
OHM
with engineering prefixes, e.g.:
MOHM
is Megaohms.
*RST value 50.0
:OUTP:IMP 50OHM
Set OUTPUT impedance to 50