Lumex Syatems LF1050 Personal Lift User Manual


 
LF1050 Patient Lift Owner’s Manual14
MAINTENANCE SCHEDULE
Item
Inspect
initially
Inspect/
adjust
daily
Inspect/
adjust
monthly
Inspect/adjust
every
6 months
Base, Casters, and Foot Pedal
Ensure that base is level (all four wheels touch the floor) x x x
Ensure that bolts and nuts are tight x x x
Ensure that casters swivel and roll freely x x x
Apply a light grease to caster ball bearings x
Check rubber parts of casters for deflection x x
Check welding joints for rust and crack x
Base Adjustments
Ensure that base opens and closes smoothly x x x x
Ensure that base locks securely in three positions x x x
Ensure that locking knob is in place and operating properly x x x
Inspect all base hardware for wear x
Mast
Ensure that mast is fully engaged and locked into mast sleeve x x
Inspect mast for bends or deflections x
Inspect all mast hardware for wear x x
Boom
Inspect boom for bends or deflections x x
Ensure that boom is centered between the base legs x x x
Inspect all boom hardware for wear x
Inspect sling hooks for wear or deflections x x
Check connections between boom and mast for improper
connection, looseness, or wear
x x
Slings
Inspect slings and straps for wear or damage x x
Inspect metal/plastic parts for wear or damage x x
Lubrication with light oil
Lubricate all pivot points x
Lubricate top and bottom of actuator mounting x
Cleaning
Clean with mild non-abrasive detergent x
Electronics and actuator
Ensure that battery is charged x x
Inspect for proper up / down operation x x
Ensure that mounting hardware is tight x x
Inspect mounting hardware for wear x
Ensure that control box is firmly affixed to the mast x x
Ensure the actuator connecting cable is tight x x
Note: Only qualified persons should service and repair your Lumex patient lift.
Regular maintenance of your patient lift is necessary to ensure continuing
proper and safe operation. Read and observe the following
recommended maintenance schedule: