ABPM 6100 Hardware Manual
DIR 80012720 Ver. B 23 / 30
7. Specifications
This section provides the unit specifications for the ABPM 6100 ambulatory blood pressure
monitor.
Power Requirements:
• Two “AA” alkaline batteries or high capacity rechargeable batteries (NimH).
Dimensions:
• 124 x 70 x 33 mm (approximate).
Weight:
• 270g (including batteries, approximate).
Operating Ambients:
• Temperature: 10°C (50°F) to 50°C (122°F).
• Humidity: 20% to 95% RH non-condensing.
Storage Ambients:
• Temperature: -20°C (-68°F) to 70°C (158°F).
• Humidity: 15% to 95% RH non-condensing.
• Altitude: -170 to 1,700 meters
Gauging Ranges:
• Systolic Blood Pressure: 60 to 250 mmHg
• Diastolic Blood Pressure: 25 to 200 mmHg
• Maximum Inflation: 270 mmHg
• Heart Rate: 40 to 200 bpm
Memory:
• Up to 250 readings using Alkaline batteries
• Up to 110 reading using rechargeable NimH
Method of Measurement:
• Oscillometric with step deflation.
Accuracy:
• ± 3 mmHg
International Standards:
• EN 60601-1-1:2001 ; Medical electrical equipment – Part 1-1: General requirements for
safety. Collateral standard: Safety requirements for medical electrical systems
Specific requirements for the protection against hazards of ignition of flammable
anesthetic mixtures and to control the risk of fire (Fire prevention)
• EN 60601-2-30:2000 ; Medical electrical equipment – Part 2-30: Particular requirements
for the safety, including essential performance, of automatic cycling non-invasive blood
pressuring monitoring equipment
• EN 60601-1-2:2001 ; Medical Electrical Equipment – Part 1-2: General requirements for
safety. Collateral Standard: Electromagnetic compatibility – requirements and tests
• EN 1060-1:1996 ; Specification for Non-invasive sphygmomanometers – Part 1.
General requirements
• EN 1060-3:1997 ; Non-invasive sphygmomanometers – Part 3: Supplementary
requirements for electro-mechanical blood pressure measuring systems
• AAMI SP 10 ES1 Category C: 1992 (battery powered) Electronic or automated
sphygmomanometers
Calibration:
• Minimally, once per year
Safety System:
• Maximum Inflation pressure limited to 300 mmHg.
• Auto safety release valve for power failure.
• Maximum BP measurement time limited to less than 180 seconds.
Data Connector:
• Stereo mini-plug headphone type connector.