Welch Allyn Medical Diagnostic Equipment ABPM 6100 Blood Pressure Monitor User Manual


 
ABPM 6100 Hardware Manual
DIR 80012720 Ver. B 22 / 30
Welch Allyn recommends using the most sensitive pressure standard possible when
performing calibration checks. A Setra Pressure Meter 2270-01, which is calibrated for
± 0.1 mm Hg, works well for this application.In the Federal Republic of Germany: In
accordance with Med BetreibV, the ABPM must be returned every 2 years to
instrumentation control for calibration certification.