Unisar BR101 Thermometer User Manual


 
NOTE: According to extensive clinical
examinations, the temperature readings
given by an ear thermometer are equivalent
to those given by its oral counterparts.
Use of this ear thermometer should not
replace a visit to your doctor.
TROUBLESHOOTING
1. Consistent low temperature readings
a. The probe is not positioned properly in
the ear canal. The tip of the probe must
be snug and fully seated against the
opening of the ear canal. Failure to
properly position the probe may lead to
a low temperature reading.
b. The waterproof lens is dirty. Clean the
lens with a soft cloth moistened with
water or alcohol.
2. The measurement is out of the range
of the ear thermometer
a. Switch the ear thermometer on before
positioning it in the ear. Please
properly position the ear thermometer
in the ear canal and then repeat the
measurement.
3. Low battery warning
a. Battery needs to be replaced.
See Battery Installation section of the
manual.
NOTE: If you believe your thermometer
is defective or otherwise not operating
properly, please check the battery first.
If the thermometer is still not operating
properly or if you are having any other
technical problems, please contact:
Unisar Inc.
15 West 36th Street
New York, NY 10018
1-888-232-6476
TECHNICAL SPECIFICATIONS
• For household use only
• Temperature readings are to the nearest
0.1°F or °C
• Temperature measurement range:
93.2° ~ 109.4°F/34°C ~ 43°C
• Ambient operating environment:
60.8 ~ 104°F (16° ~ 40°C)
• Ambient storage environment:
-4° ~ 122°F/-20° ~ 50°C
• Laboratory accuracy against an infrared
reference source:
95°~102.2°F/35° ~ 39°C: +/- 0.4°F/0.2°C
less than 96.8°F/36°C: +/- 0.5°F/0.3°C
greater than 102.2°F/39°C: +/-0.5°F/0.3°C
• Requires one 3-volt CR-2032 battery
(included)
• Equipment is not category AP/APG
12
13