Teledyne AX300-I Oxygen Equipment User Manual


 
Introduction AX300-I
Teledyne Analytical Instruments 10
The instrument is powered by three AA alkaline batteries and is
designed to operate for 2000 hours in Non-alarm State.
Oxygen analysis is linear across the single range of 0-100% using
Teledyne’s class R17MED oxygen sensor. A unique sensor failure
alarm is incorporated which warns the user if the sensor signal is lost or
low. When this occurs, the SENSOR display flashes. The alarm buzzer
can be silenced by pressing the ALARM SILENCE key () or
pressing the unlock key.
1.1 Applicable Standards
The AX300-I is built to meet or exceed regulatory and industry
standards for use as a medical device. These instruments are designed,
built and tested with the following applicable standards:
ASTM F 1462: Specifications for Oxygen Analyzers
ASTM F 1463: Specifications for Alarm Signals
ISO 7767: Oxygen Analyzers for Analyzing patient
Breathing Mixtures
ISO 9703-1: Anesthesia and Respiratory Care Alarm Signals
Part 1
ISO 9703-2: Anesthesia and Respiratory Care Alarm Signals
Part 2
EN/IEC 60601-1-2: Medical Electrical Equipment—Part 1 General
Requirements for Safety. Electromagnetic
Compatibility Requirements and Test
MIL-STD-810E: Environmental Test Methods
EN/IEC 60601-1: Medical Electrical Equipment-General
Requirements for Safety
1.2 Features
The AX300-I is a compact, versatile instrument capable of rapidly
measuring the oxygen content of an atmosphere or environment
accurately to ±2% over the range 0-100% oxygen. The following
features are standard on the AX300-I:
Large easy to read 3 1/2 digit LCD display (see options)
Automatic LCD back lighting upon key press