Connection and Installation of the Equipment
2.Preparation for Use
1. Prepare Power Source
<When using dry cell batteries>
Please use six C alkaline manganese batteries (LR14) (not
included).
Remove the lid of the Battery Compartment on the bottom of
the Scale Unit Case, place batteries correctly as instructed
inside the Battery Compartment and put the lid back on. If
the batteries are running out, the Display will show .
In that case, please change all six batteries.
ƽ
•Do not apply excessive force to the equipment when assembling or when changing batteries.
•When changing batteries, ensure that the new batteries are inserted with the correct polarity and
.
If the polarity is incorrect, then the batteries may leak and damage the equipment.
•When not in use for a long time, remove the batteries before storing the equipment.
<When using the original AC adapter>
Connect the plug of the AC Adapter to the DC jack on
the bottom of the Scale Casing and connect the AC
Adapter to the wall outlet. Then confirm the Pilot
Lamp on the AC Adapter lights up.
ƽ
Warning
•Do not insert or remove the plug with wet hands.
•Do not use the equipment near water.
ƽ
Caution
•Ensure that the adapter supplied (Model: A30930G) is
always used with this equipment.Other adapters may
cause malfunction.
2. Level Adjustment
For accurate measurement, place the unit on as flat a surface
as possible and adjust the level with the four adjustable feet
so the bubble in the level gauge is in the center of the frame.
ƽ
Caution
•Ensure the weighing platform is placed on a level and
stable surface. If the equipment is used when unstable,
for example if not all feet are on a level surface, then
there is a risk that the scale will tip over and it will be
impossible to obtain a measurement.
Adjust the height by
rotating the adjustable feet.
Plug
AC Adapter
9
Pilot Lamp
3.Operating Instructions
ƽ
• When the equipment has been transferred to another location with a temperature difference of more than
20 C, leave at least two hours before using.
• Do not use transmitters, such as mobile phones, near the equipment when in use. This may affect readings.
ƽ
• The 6kg mode has a resolution of 2g and the 15kg mode has a resolution of 5g.
• If the weight exceeds 6kg, the mode will switch to the 15kg version automatically.
1. Turn on the Power
Press the [ ] key.
will flash for four seconds on the Display.
will be displayed and an arrow will appear
next to the “STABILIZED” symbol.
2. Put the Baby into the Baby Basket
An arrow (t) will appear next to “STABILIZED” and the weight will be
displayed.
The weight will be automatically locked in when the weight of the baby
is stabilized and an arrow (
t) will appear next to "WEIGHT LOCK".
If the weight exceeds 6kg, the mode will switch to the 15kg version
automatically.
* "Weight Lock" function fluctuates and does not lock-in any numbers
when the weight of the baby is less then 500g.
To deactivate "WEIGHT LOCK" function, press "WEIGHT LOCK OFF"
key during or after measurement. "WEIGHT LOCK" function will
automatically reactivate for subsequent measurements.
3. Measurement is Completed
Remove the baby from the Baby Basket.
will be displayed and
t
will appear next to
the “STABILIZED” symbol. The mode will set at the
6kg version automatically.
ƽ
If is not displayed, press the [ ]
key.
4. Subsequent Measurements
After confirming the is displayed, place the
baby into the Baby Basket.
5. Turn off the Power
Press the [ ] key.
/
0
/