Sunrise Medical 515A Oxygen Equipment User Manual


 
SPECIFICA
TIONS
DEVILBISS 5-LITER SERIES
Catalog Number 515ADS, 515ADZ
Delivery Rate (Lower delivery rates 1 to 5 LPM
available for low flow applications)
Maximum Recommended Flo
w
5 LPM
Outlet Pressure 8.5 psig (58.6 kPa)
Electrical Rating 115 V, 60 Hz, 4.1 Amp
Oper
ating Voltage Range
97-127 V~, 60 Hz
Oxygen Percentage 1-5 LPM=93%±3%
Oper
ating Altitude
(tested at 21
°C only)
A
cross the voltage range:
0-1500 M (0-4921 ft) No degradation of performance
T
ested at nominal voltage only:
1500-3000 M (4921-9842 ft)
No degradation in performance expected based on tests of similar products.
3000-4000 M (9842-13123 ft) Not recommended/Not tested
Operating Environment Range* No degradation in performance across
10
°C to 35°C, humidity range the operating voltage range
of 30% to 75% (tested at sea level).
Po
wer Consumption
385 Watts A
verage
515ADS only-320 Watts @ 2.5 LPM & below
Weight 50 lbs. (22.7 Kilogram
s)
Sound Level (ISO 8359:1996 from front) 48.0 dbA overall average
Dimensions 27.75"H x 16"W x 14"D (70.5 x 40.6 x 35.6 cm)
Pressure Relief Valve 40 psig±5psig (276 kPa±34.5 kPa)
Operating System Time Cycle / Pressure Swing
For units equipped with an OSD 84% ± 2%
the visible “low oxygen” indicator (The audible alarm will alert at approximately 75%.
will activate at the following level At less than 60%, the red “service required” light will activate.)
Storage Conditions -40°C to 70°C, humidity range of 10% to 100%, including condensation
Equipment Class and Type Class II Equipment Double Insulated; Type B Applied Part
Approval Body CSA
and Safety Standard CAN/CSA-C22.2 No. 601.1-M90
EMC Compliance To: IEC 60601-1-2
*Note: The OSD performance at 10°C to 35°C, 95% R.H. through voltage range on the 515ADS verified at 670m.
Specifications subject to change without notice.
A-515A
Specifications
9
English