Soehnle 7700 Scale User Manual


 
21
ENGLISH
CAL
POS
R/W
CONTENTS
VALUE
RANGE
COMMENT
DEFAULT
00
R
Version number
0.00..9.99
Current version
1.00
01
W
Backlighting
In rechargable mode, on for 30 sec.
(always on with power supply)
0
1
Off
On by pressing a key or making a
weight change
0
02
W
Auto Off delay
0
1..60
No automatic switch-off
Switch off delay in minutes
3
03
W
Empty value
0..99.9
Empty value in % of max. load
1.0
04
W
Function key
0
1
2
3
4
Hold shift
Count mode
Kg/lb switch
Special dialysis function
BMI operation (medicine)
1
05
W
Hold mode
Function only > Empty Value
0
1
2
3
4
5
6
Inactive
Hold on : Remove by:
Standstill Pressing On key
Standstill Removing load
Max. value Pressing On key
Max. value Removing load
Slave counter Pressing On key
Slave counter Removing load
0
06
W
Filter setting
0
1
2
3
4
Very rapid filter
Rapid filter
Normal filter
Slow filter
Very slow filter
2
07
W
Printer
0
1
2
3
4
No printer
Strip printer
Label printer TSP800
IT
Customer-specific print image *1)
1
08
W
External alibi memory
0
1
No external alibi memory
External alibi memory available
0
09
W
Line break symbol
0
1
2
3
CR+LF
CR
LF
Semicolon
0
10
W
Decimal separator
0
1
Comma
Full stop
0
11
W
IT data set with STX-ETX frame
0
1
Inactive
STX-ETX active
0
12
W
Default function
0
1
2
Inactive
Default values, user data
Default values, print image
0