Rosewill i-Ccurate Thermometer User Manual


 
Infrared Ear Thermometers iCcurate User Manual
4
should result in a more accurate reading.
9. The thermometer is designed to take accurate temperatures with a probe cover in place. Temperatures taken without a probe cover in place.
Temperatures taken without a probe cover attached will not be accurate, and can result in damage to the soft lens. Change the probe cover after
each use to avoid infection and cross contamination.
Troubleshooting
MESSAGE SITUATION SOLUTION
E1 Operating temperature exceeds the
range of 10 - 40(50 - 104)
Move to a room within the operating range
wait 30 minutes before taking temperature
E2 Ambient temperature
changes too fast
Wait until the ambient temperature
is stable
E3 System Error Contact the retailer or service center
HI
LO
Temperature taken is not within
Typical human temperature range.
(34 - 42.2 or 93.2 - 108)
Make sure a new and clean probe
cover is attached to the thermometer
Probe. Then, take a new temperature
Battery capacity is too low. Taking
Temperature is not allowed.
Install a new battery
(1) Power is off.
(2) Improper battery installation.
(3)The battery is exhausted.
(4)Display remains blank.
(1)Press ON/M button again.
(2)Check the battery polarity.
(3)Replace with a new battery.
(4)Contact the retailer or service center.
Cleaning, Care and Storage
The probe tip is the most fragile portion of the thermometer and the soft lens is very delicate.
It is very important to always keep the probe cover attached to protect the soft lens from dirt and damage.
Whenever replacing the probe cover, inspect the soft lens and make sure it is clean and intact.
You may use cotton swab dampened with isopropyl (rubbing alcohol) or warm water to gently clean the soft lens, if dirty.
Use a clean, soft cloth to clean the surface of the device and LCD. Do not use solvents or immerse the device into water or other liquids.
Always keep the thermometer and probe covers within the storage temperature range (- 20 to 50 or – 4 to 122) and humidity range (15% -
95% non-condensing).
It is recommended to store the thermometer and probe covers in a dry location free from dust.
When not in use, store the ear thermometer inside the carry case with a new probe cover attached to the probe.
Keep the thermometer, probe covers and battery out of reach of children.
When the ambient temperature of the thermometer changes too much, such as moving the thermometer from one place of lower temperature to
another place of higher temperature, allow the thermometer to remain in a room for 30 minutes where the temperature is between10 to 40 (50 to
104).