3Using the VPAP Auto 25, VPAP ST & VPAP S
English
Supplemental Oxygen
The VPAP Auto 25, VPAP ST and VPAP S are designed to be compatible with the
following levels of supplemental oxygen usage:
Up to 4 L/min in VAuto mode
Up to 15 L/min in CPAP, S, ST and T modes.
At a fixed flow rate of supplemental oxygen flow, the inhaled oxygen concentration
will vary, depending on the pressure settings, patient breathing pattern, mask
selection, and the leak rate.
!
WARNING
Always use the ResMed Oxygen Connector Port when adding supplemental
oxygen at the flow generator outlet.
Using the VPAP Auto 25, VPAP ST & VPAP S
Setting up your VPAP
1 Connect the power cord to the socket at the rear of your VPAP.
2 Plug the other end of the power cord into the power outlet.
3 Connect one end of the air tubing firmly onto the air outlet.
4 Connect the assembled mask system to the free end of air tubing.
For information on assembling your mask, see your mask manual.
!
WARNING
Make sure the power cord and plug are in good condition and the
equipment is not damaged.
Only ResMed air tubing should be used with the device. A different type of
air tubing may alter the pressure you actually receive, reducing the
effectiveness of your treatment.
Blocking the hose and/or air inlet of the device while in operation could lead
to overheating of the device.
1
2
3
4